| Measured by | Austrian Institute of Technology | |||
|---|---|---|---|---|
| Measured with | NETZSCH DSC 204 F1 Phoenix | |||
| Measured according to standard | None | |||
| Comment on measurement | Crucible: Al. pan, closed. Purge gas: Nitrogen with 20 ml/min. 3 Heating & Cooling cycles between 140 and 220 °C with 10 K/min. Heat calibration performed. | |||
| Reference | ||||
| Heating rate | 10 K/min | |||
| Cooling rate | 10 K/min | |||
| Sample mass | 11.67 mg | |||
| Endothermic heat flow | Exothermic heat flow | |||
| Value | Uncertainty | Value | Uncertainty | |
| Enthalpy | 316.07 kJ/kg | 9.68 kJ/kg | 282.73 kJ/kg | 2.6 kJ/kg |
| Integration lower limit | 175.63 °C | 174.6 °C | ||
| Integration upper limit | 215.6 °C | 147.3 °C | ||
| Onset temperature | 185.93 °C | 0.55 K | 173.47 °C | 0.33 K |
| Endset temperature | 201.5 °C | 0.53 K | 159.57 °C | 0.24 K |