| Measured by | Austrian Institute of Technology | |||
|---|---|---|---|---|
| Measured with | NETZSCH DSC 204 F1 Phoenix | |||
| Measured according to standard | None | |||
| Comment on measurement | Crucible: Al. pan, closed. Purge gas: Nitrogen with 20 ml/min. 2 Heating & Cooling cycles between 140 and 220 °C with 10 K/min. Heat calibration performed. | |||
| Reference | ||||
| Heating rate | 10 K/min | |||
| Cooling rate | 10 K/min | |||
| Sample mass | 7.12 mg | |||
| Endothermic heat flow | Exothermic heat flow | |||
| Value | Uncertainty | Value | Uncertainty | |
| Enthalpy | 319.85 kJ/kg | 6.5 kJ/kg | 284.1 kJ/kg | 3.6 kJ/kg |
| Integration lower limit | 175 °C | 175 °C | ||
| Integration upper limit | 213.2 °C | 151.3 °C | ||
| Onset temperature | 185.5 °C | 0.2 K | 174.1 °C | 0.2 K |
| Endset temperature | 197.6 °C | 0.4 K | 163.45 °C | 0.1 K |