| Measured by | ZAE Bayern | |||
|---|---|---|---|---|
| Measured with | DSC Q2000 (TA Instruments) | |||
| Measured according to standard | Task 4229 PCM DSC procedure | |||
| Comment on measurement | Different integration limits to evaluate the crystallization enthalpy were used. | |||
| Reference | ||||
| Heating rate | 0.5 K/min | |||
| Cooling rate | 0.5 K/min | |||
| Sample mass | 10.53 mg | |||
| Endothermic heat flow | Exothermic heat flow | |||
| Value | Uncertainty | Value | Uncertainty | |
| Enthalpy | 135.2 kJ/kg | 6.8 kJ/kg | 114.6 kJ/kg | 5.7 kJ/kg |
| Integration lower limit | 25 °C | |||
| Integration upper limit | 40 °C | |||
| Onset temperature | 33.3 °C | 0.1 K | 2.3 °C | 1.6 K |
| Endset temperature | 35.2 °C | 0.1 K | 1.9 °C | 1.6 K |