| Measured by | Fraunhofer Institut | |||
|---|---|---|---|---|
| Measured with | Q2500 TA Instruments | |||
| Measured according to standard | Task 4229 PCM DSC procedure | |||
| Comment on measurement | Kommentar xyz... | |||
| Reference | ||||
| Heating rate | 0.5 K/min | |||
| Cooling rate | 0.5 K/min | |||
| Sample mass | 12.68 mg | |||
| Endothermic heat flow | Exothermic heat flow | |||
| Value | Uncertainty | Value | Uncertainty | |
| Enthalpy | 230 kJ/kg | 1 kJ/kg | 226.1 kJ/kg | 0.3 kJ/kg |
| Integration lower limit | 85 °C | 135 °C | ||
| Integration upper limit | 85 °C | 128 °C | ||
| Onset temperature | 128.68 °C | 0.06 K | 134.12 °C | 0 K |
| Endset temperature | 124.87 °C | 0.03 K | 121.87 °C | 0.02 K |